3-81
Performance Tests
SDH Frequency Offsets (Options US1, [US5], A1T, [A1U], A3R [A3S],
120)
8. Disconnect all the test equipment.
Table 3-4 Measured Offset
Offset
(ppm)
Min Expected
Frequency
Max Expected
Frequency
-100ppm MF - (0.00010452 x MF) MF - (0.00009548) x MF
-66.6ppm MF - (0.00007114 x MF) MF - (0.00006208) x MF
+33.3ppm MF + (0.00002878 x MF) MF + (0.00003782) x MF
+100ppm MF + (0.00009548 x MF) MF + (0.00010452) x MF
+999ppm MF + (0.00099448 x MF) MF + (0.00100352) x MF
-999ppm MF - (0.00100352 x MF) MF - (0.00099448) x MF