3-241
Performance Tests
Performance Test Record
+ve Pulse Width 639 ns 781 ns
Overshoot 0.474 V
Undershoot 0.474 V
704 kb/s Out 4
(UKK only)
+ve Pulse Amp 2.133 V pk 2.607 V pk
+ve Pulse Width 639 ns 781 ns
Overshoot 0.474 V
Undershoot 0.474 V
Step 8: -ve Pulse Amp 2.133 V pk 2.607 V pk
-ve Pulse Width 639 ns 781 ns
Step 8: Overshoot 0.474 V
Undershoot 0.474 V
704 kb/s Out 3
(UKK only)
Step 9: -ve Pulse Amp 2.133 V pk 2.607 V pk
-ve Pulse Width 639 ns 781 ns
Overshoot 0.474 V
Undershoot 0.474 V
704 kb/s Out 2
(UKK only)
-ve Pulse Amp 2.133 V pk 2.607 V pk
-ve Pulse Width 639 ns 781 ns
Overshoot 0.474 V
Undershoot 0.474 V
Step 12: Delay 1 to 2 4 bits
3-25 Step 14: Delay 1 to 3 8 bits
Table 11 Performance Test Record
Page
No.
Test Description Result
Min Actual Max

Table of Contents