Selecting Test Features
Adding Pointer Adjustments
Adding Pointer Adjustments
Description The transmitted SPE or VT pointer value can be adjusted for testing
purposes.
HOW TO: 1 Set up the SONET transmit interface and payload required. See
“Description” page 6.
2 Choose the POINTER TYPE.
3 Choose the ADJUSTMENT TYPE required.
BURST - You determine the size of the burst by the number of
PLACES chosen. If, for example, you choose 5 PLACES the pointer
value will be stepped 5 times in unit steps e.g. 0 (start value), 1, 2, 3,
4, 5 (final value). The interval between steps is as follows:
The SPE pointer minimum spacing between adjustments is 500 us.
For VT pointers the minimum spacing between adjustments is 2 ms.
Choose ADJUST POINTER [ON] to add the chosen burst.
NEW POINTER - You can choose a pointer value in the range 0 to 782
with or without a New Data Flag.
The current pointer value is displayed for information purposes.
Choose ADJUST POINTER [ON] to transmit the new pointer value.