8-336.  If  the Schrnitt outputs are missing or incorrect: 
a.  Verify that the correct input 
is 
present before replacing U22. If  only one of  the two ECL outputs is in- 
correct, check the resistor pack, R37, for opens or shorts before replacing U22. 
8-337.  The Comparator Offset adjustment (AVOS, 
R58) can be checked using the same setup as 
in 
the pre- 
vious tests: 
a. 
Note the present voltage at 
U22A pin 8. This voltage is needed so that the offset can be returned to 
its original value. 
b. 
Vary the Channel A voltage offset resistor (AVOS, 
R58) over its 
full 
range. The voltage at pin 8 
should vary from 
+ 
20 
to 
- 
20 
millivolts. 
c.  If  this voltage is incorrect, check the 
+ 
5 and 
- 
5.2 supply voltages at R58. Also check the divider 
resistors R56 and R57 
as 
well as R58 for opens or shorts. 
d.  Remember to reset the voltage at pin 8 to its original value after making this check. 
8-338.  If  the SENS control does not work when the Counter is in the Sensitivity mode: 
a. 
Check the LATCH ENABLE 
lie at pin 
4 
of  U22A. The voltage should vary from approximately 
+50 to -50  millivolts as the SENS control is adjusted over its 
full 
range. If  these voltages are incor- 
rect, check the SENS control or the DAC block switching circuitry. If  the voltage has the correct 
range, U22 may be defective. 
8-339.  In the last part of  the Channel A circuit, the BUFFER STAGE, a problem can be partially isolated by 
observing the failure symptoms of  the Counter. During a measurement, the display may be incorrect or missing, 
or the front panel trigger light may fail to flash. If  only one of  these failures occur, the fault is probably in the 
buffer stage. Replacing U21 may 
fur 
the problem. However, 
if 
both of  these symptoms occur together, verify the 
output of  the previous Schmitt stage before troubleshooting the buffer stage. 
8-340.  To check the Buffer Stage: 
Settings of 
5334B 
Settings of  Signal Source 
FREQ A 
....................................... 
ON 
1 
MHz, 
1 
Volt p-p sine wave 
(set 
HP3325A 
AMPTD 
to 
.5V 
to generate a 
1V 
p-p 
signal) 
0 volt, dc offset 
50R 
.................................................... 
ON 
AUTO TRIG 
................................... 
OFF 
SENS 
................................................. 
ON 
SENS control 
................................. 
Fully 
CW 
a.  Apply the signal to the Channel 
A 
Input and note the failure symptoms. 
b. 
A display problem is caused by  the Level-Shifter, U21C which provides the input to the Multiple- 
Register Counter (MRC). 
A 
defective Trigger Light function is caused by  the Trigger Light One- 
Shot, U21B. 
c. 
Measure the output of  the level-shifter 
U21C, at pin  12. If  the 
1 
MHz, TTL signd  is missing or incor- 
rect, verify the input to the level-shifter at pins 10 and 
11. 
d.  If  the complementary ECL signals are not present, check the damping and pull-down resistor packs, 
R36 and R37, for opens or shorts. Verify that the output of  the Schmitt Stage is correct. 
HP 