NOTE
For
53348
instruments equipped with Option
060,
set Signal Generator
amplitude to
40.0
mV
k
0.4
mV.
16.
Verify that the 5334B front panel displays 70.000 000
MHz
a
4
Hz
at both the positive and negative slope
settings.
17.
Set the Signal Generator frequency to 104
MHz
and the amplitude to 28.0 mV
a
0.3 mV
NOTE
For
53348
instruments equipped with Option
060,
set Signal Generator
amplitude to
40.0
mV
st
0.4
mV.
18.
Verify that the 5334B front panel displays 104.000 000 MHz
k
4
Hz
at both the positive and negativc slope
settings.
5-19.
MULTIPLE-REGISTER COUNTER (MRC) INPUT ADJUSTMENT, CH
B
Reference:
Figztre
8-20
Description:
The Channel
B
DC biasing level on the MRC input signal is optimized at two frcqucncics on
bo~h
the positive and negativc slopes of the signal (interactive adjustment).
r
\
HP8656B
SIGNAL GENERATOR
~'I*E'A~'
IN
HP5334B
UNIVERSAL COUNTER
TIMEBASE
OUT
2
7523-L2
Figire
5-9.
MRC Ittpul Adj~tsllCltartrtel
B
Sehip
HP
5334B
-
Service Manual
5-15