NOTE 
For 
53348 
instruments equipped  with  Option 
060, 
set  Signal  Generator 
amplitude to 
40.0 
mV 
+ 
0.4 
mV.  Terminate the  unused Channel 
A 
and 
B 
rear inputs with 
50R 
loads. 
2. 
Set the 
5334B 
as follows: 
FUNCTION 
................................ 
FREQ A 
SENS (blue key) 
.............................. 
ON 
INPUT A SENS Control 
............... 
MAX 
INPUT A 
50n 
................................. 
ON 
GATE TIME 
................................... 
1 
Second 
INPUT A SLOPE 
........................... 
OFF (positive) 
3. 
Connect the Signal Generator output to the 
5334B 
Input A. Connect Time Base as shown in 
Figztre 
5-8. 
4. 
Adjust 
AIR130 
(CH A) for a 
5334B 
front panel display of 
70.000 000 
MHz 
+ 
4 
Hz at both the positive and 
negative slope settings. 
5. 
Decrement the Signal Generator amplitude by 
3 
mV  rms. 
6. 
Adjust 
AIR130 
(CH A) for a 
5334B 
front panel display of 
70.000 
000 
MHz 
+ 
4 
Hz at both the positive and 
negative slope settings. If  this adjustment is not possible because the signal level has dropped below the 
Counter's sensitivity limit, go to step 
10. 
Otherwise, continue on to step 
7. 
7. 
Set the Signal Generator frequency to 
104 
MHz. 
8. 
Adjust 
AIR130 
(CH A) for a 
5334B 
front panel display of 
104.000 000 
MHz 
+- 
4 
Hz at both the positive 
and negative slope settings. If  this adjustment is not possible, go to step 
10. 
Otherwise continue on to step 
9. 
9. 
Set the Signal Generator frequency to 
70 
MHz and repeat steps 
5 
through 
9 
until adjustment is no longer 
possible and this loop is exited from step 
6 
or 
8. 
10. 
Increment the Signal Generator amplitude in 
0.3 
mV 
rms steps until Counter displays a stable frequency 
count. 
11. 
Set the Signal Generator frequency to 
104 
MHz. 
12. 
Adjust 
AIR130 
(CH A) for a 
5334B 
front panel display of 
104.000 000 
MHz 
t 
4 
Hz at both the positive 
and negative slope trigger settings. If  this adjustment is  not possible, repeat steps 
10 
through 
12 
until this 
ad- 
justment is completed and then continue 
with 
step 
13. 
13. 
Set the Signal Generator frequency to 
70 
MHz. 
14.  Adjust 
AIR130 (CH 
A) 
for a 
5334B 
front panel display of 
70.000 
000 
MHz 
k 
4 
Hz at both the positive and 
negative slope settings. If  this adjustment is not possible, repeat steps 
10 
through 
14 
until this adjustment is 
completed and then continue with step 
15. 
15. 
Set the Signal Generator frequency to 
70 
MHz and the amplitude to 
28.0 
mV 
+. 
0.3 
mV. 
HP 
5334B 
- 
Service Manual 
5-14