Procedure:
1.
Allow the ovenized oscillator a 24-hour warm-up period prior to performing this adjustment. The crystal os-
cillator oven is operating whenever the instrument is connected to its power source.
2.
Set the oscilloscope as follows:
...........................
Channel A VPiv. 0.5V
TimePiv
...........................................
0.01~s
......................
Ch
A
Input Coupling 50fl
Main Triggering
..............................
EXT
...............................
Vertical Display A
..........................
Horizontal Display MAIN
3.
Set the 5334B as follows:
.....................................
FUNCTION
FREQ
A
GATE TIME
...................................
1
Second
TIME BASE
....................................
INT (rear panel)
4. Connect the 10 MHz Frequency Standard to the oscilloscope External Trigger input.
5. Connect the TIME BASE output to the oscilloscope
channel A input.
6.
Adjust the "FREQ ADJ" pot on the Option 010 Oven Oscillator until the scope pattern does not move
sideways more than
1
division in 10 seconds.
7.
Set the oscilloscope as follows:
..............................
Main Triggering INT
...............................
Internal Trigger A
--
............
Channel A Input Coupling 500
8.
Connect the 5334~ TIME BASE output to the oscilloscope channel A input.
a
9.
Verify that the signal is greater than 1.4V p-p.
5-16.
CHANNEL A SENSITIVITY OFFSET ADJUSTMENT
Reference: Figure
8-20
and
8-21
Description:
The sensitivity of the Channel A Input is adjusted using a
1
MHz input signal (interactive adjust-
ment).
HP
5334B
-
Service Manual
5-9