4.
Adjust AlCl (OSC
ADJ)
for a 5334B front panel display of 10.000
000
MHz
2
8
Hz.
5.
Set the oscilloscope as follows:
Channel
A
V/Div.
...........................
0.5V
Time/Div
...........................................
0.05ps
Ch A Input Coupling
......................
5051
Main
Triggering
..............................
INT
Internal Trigger
...............................
A
Vertical Display
...............................
A
Horizontal Display
..........................
MAIN
6.
Connect the 5334B TIME BASE output to the oscilloscope channel A input.
7.
Verify that the signal is greater than 1.4V p-p.
NOTE
The following procedure is for the Option
010
Oven Oscillator equipped
53348.
2
HP1715A
OR
EQUIVALENT
OSCILLOSCOPE
TSlB-L2
Figure
5-5.
Option
010
Oscillator Adjustment Setup
Equipment:
Oscilloscope
.....................................
HP
1715A (or equivalent)
Ref. Freq. Std.
................................. 10
MHz
Reference:
Figure
8-27
C
HP
5334B
-
Service
Manual
5-8