2.
Set the HP 5334B as follows:
Reinitialize the 5334B.
FVNCTION
...........................................................................................................
FREQ B
SENS
......................................................................................................................
ON
...........................................................................
CHAN
B TRIGISENS control
fully
CW
3.
Connect the signal generator to the HP
5334B Input B using a 50R feedthrough connector as shown in
Figure
45.
4.
Repeat the tests of paragraph 4-33 for Channel
B
and record the measurements on the Performance Test
Record Card, lines
19
through 24. Begin the tests at the verification of 80 MHz in paragraph 4-33, step 4.
Failure:
If any of these tests fail, refer to Section
V,
Adjustments, paragraphs 5-17 and 5-19 as a first step in
troubleshooting.
4-32.
CHANNEL A FREQUENCY RESPONSE AND SENSITIVITY TEST,
80 MHz-100 MHz
This test is for instrument that contain the
Option
060
Rear Panel Inputs,
i.e., instruments with both Front and
Rear Inputs.
Specification:
Refer to
Table
1-1,
HP Model 5334B Specifications, for Option
060
Frequency Response and Sen-
sitivity specifications.
Description:
The frequency measuring range of the Counter is tested at minimum sensitivity specifications and
two different frequency settings.
•
Frequencies and conditions tested:
80 MHz and 100 MHz, dc coupled,
1
MR
80 MHz and 100 MHz, ac coupled,
1
MR
REAR
A
B
n
n
I
REF
OUT
HP8660A/C/D
SIGNAL GENERATOR
+REF
IN
UNIVERSAL COUNTER
u
1
lses
~21
Figure
46.
Channel
A
Frequency and Sensitivily Sehip
for Option
060
HP 5334B
-
Service Manual
4-21