HP 5334B - 4-32. CHANNEL A FREQUENCY RESPONSE AND SENSITIVITY TEST, 80 MHz-100 MHz

HP 5334B
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.
Set the HP 5334B as follows:
Reinitialize the 5334B.
FVNCTION
...........................................................................................................
FREQ B
SENS
......................................................................................................................
ON
...........................................................................
CHAN
B TRIGISENS control
fully
CW
3.
Connect the signal generator to the HP
5334B Input B using a 50R feedthrough connector as shown in
Figure
45.
4.
Repeat the tests of paragraph 4-33 for Channel
B
and record the measurements on the Performance Test
Record Card, lines
19
through 24. Begin the tests at the verification of 80 MHz in paragraph 4-33, step 4.
Failure:
If any of these tests fail, refer to Section
V,
Adjustments, paragraphs 5-17 and 5-19 as a first step in
troubleshooting.
4-32.
CHANNEL A FREQUENCY RESPONSE AND SENSITIVITY TEST,
80 MHz-100 MHz
This test is for instrument that contain the
Option
060
Rear Panel Inputs,
i.e., instruments with both Front and
Rear Inputs.
Specification:
Refer to
Table
1-1,
HP Model 5334B Specifications, for Option
060
Frequency Response and Sen-
sitivity specifications.
Description:
The frequency measuring range of the Counter is tested at minimum sensitivity specifications and
two different frequency settings.
Frequencies and conditions tested:
80 MHz and 100 MHz, dc coupled,
1
MR
80 MHz and 100 MHz, ac coupled,
1
MR
REAR
A
B
n
n
I
REF
OUT
HP8660A/C/D
SIGNAL GENERATOR
+REF
IN
UNIVERSAL COUNTER
u
1
lses
~21
Figure
46.
Channel
A
Frequency and Sensitivily Sehip
for Option
060
HP 5334B
-
Service Manual
4-21