HP 5334B - 4-34. PERIOD A TEST

HP 5334B
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.
Set the HP
5334B
as follows:
Reinitialize the
5334B.
FUNCTION
...........................................................................................................
RE
B
SENS
.......................................................................................................................
ON
...........................................................................
CHAN
B
TRIGISENS control fully CW
3.
Connect
50n
feedthroughs or terminations on the rear panel A and
B
Inputs.
4.
Connect the signal generator to the HP
5334B
front panel Input
B
as shown in
Figure
4-7.
5.
Repeat the tests of paragraph
4-35
for Channel
B
and record the measurements on the Performance Test
Record Card, lines
29
through
32.
Begin the tests at the verification of
80
MHz
in
paragraph 4-35, step
5.
Failures: If
any
of these tests fail, refer to Section
V,
Adjustments, paragraph
5-17
and
5-19
as a first step in
troubleshooting.
4-34.
PERIOD
A
TEST
Specification: Refer to
Table
1-1,
HP Model
5334B
Specifications, for Period
A
specifications.
Description: The minimum specified period measurement of
10
ns is verified using
a
100
MHz input signal.
50n
FEEDTHROUGH
HP8660A/C/D
SIGNAL GENERATOR
1
REF
IN
I
TW-LPJ
Figure
4-8.
Period
A
Test
Setup
Equipment:
Signal Generator
...................................................................................................
HP 8660NC
HP
5334B
-
Service Manual
4-24