2.
Set the HP
5334B
as follows:
Reinitialize the
5334B.
FUNCTION
...........................................................................................................
RE
B
SENS
.......................................................................................................................
ON
...........................................................................
CHAN
B
TRIGISENS control fully CW
3.
Connect
50n
feedthroughs or terminations on the rear panel A and
B
Inputs.
4.
Connect the signal generator to the HP
5334B
front panel Input
B
as shown in
Figure
4-7.
5.
Repeat the tests of paragraph
4-35
for Channel
B
and record the measurements on the Performance Test
Record Card, lines
29
through
32.
Begin the tests at the verification of
80
MHz
in
paragraph 4-35, step
5.
Failures: If
any
of these tests fail, refer to Section
V,
Adjustments, paragraph
5-17
and
5-19
as a first step in
troubleshooting.
4-34.
PERIOD
A
TEST
Specification: Refer to
Table
1-1,
HP Model
5334B
Specifications, for Period
A
specifications.
Description: The minimum specified period measurement of
10
ns is verified using
a
100
MHz input signal.
50n
FEEDTHROUGH
HP8660A/C/D
SIGNAL GENERATOR
1
REF
IN
I
TW-LPJ
Figure
4-8.
Period
A
Test
Setup
Equipment:
Signal Generator
...................................................................................................
HP 8660NC
HP
5334B
-
Service Manual
4-24