Procedure:
1.
Set the signal generator as follows:
...............................................................................................................
Frequency 100 MHz
..............................................................................................................
Amplitude 50
mV
rms
NOTE
OPTION
060
If the
HP
53348
has Option
060
(rear panel inputs), terminate the unused
Channel
A
input (front or rear) with a
500
load.
2.
Set the
HP
5334B as follows:
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
PERIOD
A
GATE TIME
.........................................................................................................
1
Second
SENS
.......................................................................................................................
ON
CHAN
A
TRIGISENS control
...........................................................................
fully CW
3.
Connect the signal generator output to the
HP
5334B Input
A
using a 50 Ohm feedthrough connector as
shown
in
Figure
4-8.
NOTE
Do not use a
50R
feedthrough connector at the Input for Counters
with
Op-
tion
060.
VERIFY:
The Counter displays 10 ns
+.
.000 000
1
ns.
4.
Record the Period
A
measurement on the Performance Test Record Card, line 33.
Failure:
If the instrument under test does not meet the test specification, consider performing the adjustments
in
Section
V
of this manual as a first step in correcting the problem.
HP
5334B
-
Service Manual
4-25