Loading...
Equipment:
/
Figure
5-1.
+
3
Volt
Power
Supply
Adjustment Setup
AIR202
TP
(GND)
\
HP3468A
DIGITAL VOLTMETER
Digital Multitimeter (DMM)
........
HP 3468A
HP5334B
UNIVERSAL COUNTER
Procedure:
i
0
oooopoo
0000000
TS15-L2
1.
Set the DMM as follows:
Function
...........................................
Vdc
Range
................................................
AUTO
2.
Connect DMM positive lead to
TP5
(
+
3), and connect the nega-
tive lead to
"GND"
Test Point.
3.
Adjust
AIR202 for a DMM reading of +3.00V dc
2
0.02V dc.
5-13.
DIGITAL-TO-ANALOG CONVERTER (DAC) A/B ZERO ADJUSTMENT
Reference:
Figure
8-21
Description:
The zero reference positions for Channel A and
B
DAC trigger levels
are
set with no input signal.
HP
5334B
-
Service
Manual