2. 
Set the HP 
5334B 
as follows: 
Reinitialize the HP 
5334B. 
FUNCTION 
........................................................................................................... 
T.I. A-B  DELAY 
GATE TIME DELAY 
........................................................................................ 
9 
ms 
SENS 
....................................................................................................................... 
ON 
A 
& 
B 
TRIGISENS controls 
............................................................................... 
fully CW 
COM A 
................................................................................................................... 
ON 
CHAN A 
& 
B 50a 
............................................................................................... 
ON 
CHAN 
B 
Negative SLOPE 
................................................................................. 
ON (falling edge) 
3. 
Connect the function-generator output to the HP 
5334B 
Input A. 
4. 
Press SINGLE CYCLE on the HP 
5334B. 
VERIFY: 
The Counter displays 
15 
ms 
+ 
100 
ps 
5. 
Record the Time Interval Delay measurement on the Performance Test Record, line 
36. 
Failure: 
If  the instrument under test does not meet the test specification, consider performing the adjustments 
in 
Section 
V 
of  this manual as a first step in correcting the problem. 
4-38. 
RATIO A/B TEST 
Specification: 
Refer to 
Table 
1-1, 
HP 
Model 
5334B 
Specifications, for Ratio 
A/B 
specification. 
Description: 
Two different frequencies are applied to the A and 
B 
inputs. The ratio of  the A and B inputs will 
be displayed. 
FUNCT 
ION 
GENERATOR 
TSlP-LZ 
Figire 
4-12. 
Ratio 
AIB 
Test Sehip 
HP 
5334B 
- 
Service 
Manual 
4-30