HP 5334B - 4-38. RATIO A;B TEST

HP 5334B
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.
Set the HP
5334B
as follows:
Reinitialize the HP
5334B.
FUNCTION
...........................................................................................................
T.I. A-B DELAY
GATE TIME DELAY
........................................................................................
9
ms
SENS
.......................................................................................................................
ON
A
&
B
TRIGISENS controls
...............................................................................
fully CW
COM A
...................................................................................................................
ON
CHAN A
&
B 50a
...............................................................................................
ON
CHAN
B
Negative SLOPE
.................................................................................
ON (falling edge)
3.
Connect the function-generator output to the HP
5334B
Input A.
4.
Press SINGLE CYCLE on the HP
5334B.
VERIFY:
The Counter displays
15
ms
+
100
ps
5.
Record the Time Interval Delay measurement on the Performance Test Record, line
36.
Failure:
If the instrument under test does not meet the test specification, consider performing the adjustments
in
Section
V
of this manual as a first step in correcting the problem.
4-38.
RATIO A/B TEST
Specification:
Refer to
Table
1-1,
HP
Model
5334B
Specifications, for Ratio
A/B
specification.
Description:
Two different frequencies are applied to the A and
B
inputs. The ratio of the A and B inputs will
be displayed.
FUNCT
ION
GENERATOR
TSlP-LZ
Figire
4-12.
Ratio
AIB
Test Sehip
HP
5334B
-
Service
Manual
4-30