4-36.
TIME INTERVAL A
TO
B
TEST
Specification: Refer to
Table
1-1,
HP
Model
5334B
Specifications, for Time Interval
A
to
B
specification.
Description:
Time Interval measuring accuracy is verified using a known generated signal.
HP3325A
SYNTHESIZER/
FUNCTION
GENERATOR
.
1
REF
IN
HP5334B
UNIVERSAL COUNTER
I
REF
OUT
8.8
a.
ooooo
017
ooooo
[:=m
I
TS10J2
-
--
-
- -
Figure
4-10.
Time Interval
A
to3~est Setup
Equipment:
Function Generator
..............................................................................................
HP
3325A
Procedure:
1.
Set the function generator as follows:
Frequency
...............................................................................................................
5
MHz
Amplitude
..............................................................................................................
2
mV
p-p
.................................................................................................................
Function Square Wave
2.
Set the
HP
5334B
as follows:
Reinitialize the
5334B.
FUNCTION
...........................................................................................................
TI.
A+
B
COM
A
...................................................................................................................
ON
100
GATE AVERAGE
.......................................................................................
ON
SENS
.......................................................................................................................
ON
A
&
B
TRIGISENS controls
...............................................................................
fully
CW
CHAN
A
and
B
50n
..................................
:
.........................................................
ON
.................................................................................
CHAN
B
Negative SLOPE
ON (falling edge)
3.
Connect the function generator output to the
HP
5334B
Input
A.
VERIFY:
The Counter displays
100
ns
+
6
ns.
HP
5334B
-
Service Manual
4-28