HP 5334B - 4-23. CHANNEL C (Option 030)

HP 5334B
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.
Rotate Channel B trigger level control.
Observe:
There is no effect on the Counter's operation.
7.
Set AUTO TRIG to OFF. Rotate Channel
B
trigger level control.
Observe:
Extreme clockwise and counterclockwise control settings will stop the gating and update of the
Counter. Trigger light stops flashing.
Test Record:
Mark Pass or Fail on the Operational Verification Test Record Card, line
7.
What Checked:
Control lines and
DAC
circuitry.
For Failures:
Refer to Section VIII of this manual. The test involves all of the functional blocks: the Input
Amplifier, DAC, Measurement, Executive, Front Panel, and Time
BaseIPower Supply blocks.
4-23.
CHANNEL
C
(Option
030)
This operational check is for HP 5334B7s containing
Option
030.
Description:
The Channel C option is checked by simply measuring a frequency within its range of 90 MHz to
1300
MHz.
Equipment:
A signal source capable of outputting some frequency from 90 MHz to 1300 MHz.
Procedure:
1.
Set the signal source as follows:
Frequency
...............................................................................................................
90 to 1300 MHz
Amplitude
..............................................................................................................
300 mV rms
(
+
2.5 dBm)
2.
Set the
HP
5334B
as
follows:
Reinitialize the 5334B.
...........................................................................................................
FUNCTION FREQ
C
3. Connect the signal source to the
HP
5334B Input C.
4.
Adjust the Channel
C
Sensitivity control as needed to cause the Counter to gate and display a stable reading.
Observe:
The HP 5334B front panel displays the generated frequency.
Test Record:
Mark Pass or Fail on the Operational Verification Test Record Card, line
11.
What Checked:
Basic operation of the Channel C option.
For Failures:
Refer to Section VIII of this manual. The Channel C, Measurement, Executive, and Front Panel
blocks are involved here.
HP
5334B
-
Service Manual
4-10