Equipment:
Function Generator
..............................................................................................
HP
332%
Procedure:
1.
Set the function generator as follows:
..........................................................
...................................................
Frequency
10
Hz
Amplitude
..............................................................................................................
15
mV rms
Function
.................................................................................................................
Sine Wave
2.
Set the HP
5334B
as follows:
Reinitialize the
5334B.
FUNCTION
...........................................................................................................
FREQ
B
SENS
.......................................................................................................................
ON
...........................................................................
CHAN
B
TRIGISENS control
fully
CW
3.
Connect the function generator signal to the
HP
5334B
Input
B
using a
50
Ohm feedthrough connector as
shown in
Figure
4-3.
4.
Repeat the tests of paragraph
4-31
for Channel
B
and record the measurements on the Performance Test
Record Card, lines
7
through
12.
Begin the tests at the verification of
10
Hz in paragraph
4-31,
step
4.
Failure:
If any of these tests fail, refer to Section
V,
Adjustments, paragraphs
5-17
and
5-19
as a first step in
troubleshooting.
e
4-30. CHANNEL A FREQUENCY RESPONSE AND SENSITIVITY TEST,
80 MHz-100
MHz
This test is for instruments that
do
not
contain the
Option
060
Rear Panel Inputs,
i.e., instruments with Front In.
puts only.
Specification:
Refer to
Table
1-1,
HP
Model
5334B
Specifications, for Frequency Response and Sensitivity
specifications.
Description:
The frequency measuring range of the Counter is tested at minimum sensitivity specifications and
two different frequency settings.
Frequencies and conditions tested:
80
MH7 and
100
MHz, dc coupled,
1
MR
80
MHz and
100
MHz, ac coupled,
1
MR
80
MHz and
100
MHz, ac coupled,
50R
HP
5334B
-
Service Manual
4-17