HP 5334B - 4-35. PULSE WIDTH A TEST

HP 5334B
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-35.
PULSE
WIDTH
A
TEST
Specification:
Refer to
Table
1-1,
HP Model
5334B
Specifications, for Pulse Width
A
specifications.
Description:
A
pulse width
is
generated and then measured with the HP
5334B
to verify the Counter's perfor-
mance.
I
Figure
4-9.
Pulse
Width
A
Test Setup
P
\
Equipment:
Signal Generator
...................................................................................................
HP 8660AIC
HP5334B
UNIVERSAL COUNTER
Procedure:
1.
Set the signal generator as follows:
Frequency
...............................................................................................................
100
MHz
Amplitude
..............................................................................................................
200
mV
rms
Q
rr
J
HP8660A/C/D
SIGNAL GENERATOR
2. Set the HP 5334B as follows:
REF
OUT
Reinitialize the
5334B.
FUNCTION
...........................................................................................................
PULSE WIDTH
A
'
HP
5334B
-
Se~ce Manual
4-26