4-35.
PULSE
WIDTH
A
TEST
Specification:
Refer to
Table
1-1,
HP Model
5334B
Specifications, for Pulse Width
A
specifications.
Description:
A
pulse width
is
generated and then measured with the HP
5334B
to verify the Counter's perfor-
mance.
I
Figure
4-9.
Pulse
Width
A
Test Setup
P
\
Equipment:
Signal Generator
...................................................................................................
HP 8660AIC
HP5334B
UNIVERSAL COUNTER
Procedure:
1.
Set the signal generator as follows:
Frequency
...............................................................................................................
100
MHz
Amplitude
..............................................................................................................
200
mV
rms
Q
rr
J
HP8660A/C/D
SIGNAL GENERATOR
2. Set the HP 5334B as follows:
REF
OUT
Reinitialize the
5334B.
FUNCTION
...........................................................................................................
PULSE WIDTH
A
'
HP
5334B
-
Se~ce Manual
4-26