Equipment:
..............................................................................................
Function Generator HP
3325A
Procedure:
1.
Set the function generator as follows:
...............................................................................................................
Frequency
5
MHz
..............................................................................................................
Amplitude
100
mV rms
Function
.................................................................................................................Sine
Wave
2.
Set the HP
5334B
as
follows:
Reinitialize the HP
5334B.
FUNCTION
...........................................................................................................
RATIO
A/B
SENS
.......................................................................................................................
ON
A
&
B
TRIGISENS controls
...............................................................................
fully CW
CHAN
A
&
B
...............................................................................................ON
3.
Connect HP
5334B
TIME
BASE
signal to Input
A
and connect the function generator signal to Input
B.
VERIFY:
The Counter displays
2.000
000
+
000 001.
4.
Record the Ratio
A/B
measurement on the Test Record Card, line
37.
Failure: If the instrument under test does not meet the test specification, consider performing the adjustments in
Section
V
of this manual as
a
first step in correcting the problem.
HP
5334B
-
Service Manual
4-31