12.
Record the measurement on the Performance Test Record Card, line
28.
Failure:
If any of these tests fail, refer to Section
V,
Adjustments, paragraphs 5-16 and 5-18 as a first step in
troubleshooting.
4-33.
CHANNEL
B
FREQUENCY RESPONSE AND SENSITIVITY TEST,
80 MHz-100
MHz
This test is for instruments that contain the
Option
060,
i.e., instruments with both Front and Rear Inputs.
Specification:
Refer to
Table
1-1,
HP Model 5334B Specifications, for Option
060
Frequency Response and Sen-
sitivity specifications.
Description:
The frequency measuring range of the Counter is tested at minimum sensitivity specifications and
two different frequency settings.
Frequencies and conditions tested:
80
MHz and
100
MHz, dc coupled,
1
MR
80 MHz and
100
MHz, ac coupled,
1
MR
Figure
4-7.
Channel
B
Frequency
and
Sensitivity Sehrp
for Option
060
r
Equipment:
Signal Generator
..................................................................................
................
HP
8660AlC
REAR
A
B
n
n
Procedure:
1.
Set the signal generator as follows:
Frequency
..............................................
................................................................
80
MHz
Amplitude
..............................................................................................................
50 mV rms
HP8660A/C/D
SIGNAL GENERATOR
REF
IN
HP 5334B
-
Service Manual
4-23
'
REF
OUT
lqgJJJJ
-==
0
HP5334B
UNIVERSAL COUNTER
Ei-
r
i
#
TS87-L2