Equipment:
...................................................................................................
Signal Generator HP 8660AlC
Procedure:
1.
Set the signal generator as follows:
Frequency
...............................................................................................................
80 MHz
Amplitude
..............................................................................................................
50 mV rms
2.
Set the HP 5334B as follows:
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
FREQ A
SENS
.......................................................................................................................
ON
...........................................................................
CHAN
A TRIGISENS control
fully
CW
3.
Connect 50R feedthroughs or terminations on the rear panel A and B Inputs.
4.
Connect the signal generator to the HP
5334B
front panel Input A as shown in Figure
4-6.
VERIFY:
The Counter displays 80 MHz
k
2
Hz.
5.
Record the measurement on the Performance Test Record Card, line
25.
6.
Set the signal generator as follows:
...............................................................................................................
Frequency 100 MHz
..............................................................................................................
Amplitude 0 mV rms
VERIFY:
The Counter displays 100 MHz
+
2
HZ.
7.
Record the measurement on the Performance Test Record Card, line
26.
8.
Set the signal generator as follows:
...............................................................................................................
Frequency 80 MHz
..............................................................................................................
Amplitude 50 mV rms
9.
Set the HP
5334B
as follows:
AC
...........................................................................................................................
ON
VERIFY:
The Counter displays
80
MHz
+
2
Hz.
10. Record the measurement on the Performance Test Record Card, line
27.
11.
Set the signal generator as follows:
...............................................................................................................
Frequency 100 MHz
..............................................................................................................
Amplitude 50 mV rms
VERIFY:
The Counter displays 100 MHz
+
2
Hz.
HP
5334B
-
Service Manual