Page #45 background imageLoading...
Page #45 background image
400 Series
EN - 43
9. By using the P-PE mode, if
the instrument detects that,
in case the test should be
performed, a contact voltage
higher than the set limit
would build up in the system
being tested, it does not
perform the test and displays
the message reported here
to the side. Check the
LOOP
----
---A
FRQ=50.0Hz
VP-N=231V VP-Pe=231V
Ut > Ulim
Dangerous contact voltage
P-PE STD
Func Mod.
efficiency of the protective conductor and the earthing
10. If, after repeated tests, the
instrument has overheated,
the message reported here
to the side is displayed. Wait
for this message to
disappear before performing
other tests
LOOP
----
---A
FRQ=50.0Hz
VP-N=231V VP-Pe=231V
Hot temperature
Overheated instrument
P-N STD
Func Mod.
11.
The previous anomalous results cannot be saved

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HT 400 Series and is the answer not in the manual?

HT 400 Series Specifications

General IconGeneral
TypeDigital Multimeter
Diode TestYes
Auto RangingYes
Data HoldYes
Operating Temperature0°C to 40°C
DC Current Range400μA to 10A
AC Current Range400μA to 10A
Resistance Range400Ω to 40MΩ
Continuity TestYes
DisplayLCD, 4000 counts
Measurement RangeAs per individual functions

Related product manuals