3.4.6 Set-up Menu
To enter SETUP Mode follow the following procedure (also refer to the chart on page 20)
1. Press MAIN MENU to enter the Main Menu list from Normal display mode
2. Use Parameter UP & DOWN buttons to reach SETUP
3. Press ENTER
4. Enter Level 2 code
5. Use Parameter UP & DOWN buttons to reach required SETUP item
6. Press ENTER
7. Use value UP & DOWN buttons followed by ENTER to update and move to next parameter If the value is left unchanged,
ENTER will have no action.
Press MAIN MENU to get back to MAIN MENU; Press again to get back to Normal display
The SETUP menu has the following items:
Parameter Value Defaults
REMPANEL 0 or 1 0 Set to 1 to operate the display panel remotely.
REMPOD 0 or 1 0 Set to 1 if an Enviropod is connected.
DETFLOW 0 or 1 1 In addition to individual sector flow monitoring, detector flow monitoring is
also incorporated; Set to 1 to enable.
CCODE List of supported Languages See Below for table. Change metric/imperial units.
RES-ISOL 0 or 1 1 Set to 1 to reset system with 24 V on reset line.
Set to 0 to isolate relays with 24 V on reset line.
MOD1 to 5 List of available I/O modules All I/O modules fitted at build will have this information entered. If
additional I/O modules are installed, their type must be entered here.
The e-series range can operate in a number of different languages. To change the language the correct COUNTRY CODE must be entered.
The software can also be made to swap from metric to imperial measurements. This is achieved by entering the United States country
code (1), the language used is English. These are as follows:
Language Code
English 44
United States (English) 1
French 33
Spanish 34
Portuguese 35
Italian 39
German 49
The default code is 44