EasyManua.ls Logo

Idexx VetTest - Page 93

Idexx VetTest
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
*o
4FMG)FMQ(VJEF
Temperature Warnings
Problem: The VetTest analyzer takes more than 25 minutes to warm up.
Cause: The VetTest analyzer cannot reach the proper operating temperature due
to a cold or hot room, or analyzer temperature. The room temperature
TIPVMECF×o×$×o×'
Note:5IFBOBMZ[FSDIFDLTUIFUFNQFSBUVSFBUNJOVUFTTFDPOET
If the temperature is not correct, the VetTest analyzer will change the
EJTQMBZFEDPVOUEPXOUJNFUPNJOVUFTTFDPOETBOETUBZJOUIJT
warm–up cycle until it reaches the proper temperature.
Solution: Allow the room and/or the analyzer to come to proper temperature.
Very low temperatures can cause considerable damage to the
VetTest analyzer. Permanent damage may be caused if the analyzer
is switched on at temperatures below freezing. The analyzer should
CFVOQBDLFEBOEBMMPXFEUPDPNFUPSPPNUFNQFSBUVSF×o×$
×o×'GPSBNJOJNVNPGPOFIPVSBGUFSJUIBTCFFOJOBDPME
environment.
Cause: The VetTest analyzer is above proper operating temperature.
Solution: Make sure the VetTest analyzer has adequate airflow around the unit
(review Choosing the Analyzer LocationQBHF#
Cause: Electrical interference.
Solution: Try a different circuit. A surge protector, line conditioner, or dedicated
circuit may be required.
Cause: Radio frequency (RF) or electromagnetic interference (review Precautions,
QBHF#
Solution: Contact IDEXX Technical Support for assistance.
Problem: High temperature warning.
Cause: The VetTest analyzer is above proper operating temperature.
Solution: Review Choosing the Analyzer LocationQBHF#
Problem: Low temperature warning.
Cause: The VetTest analyzer warmup has been bypassed.
Solution: Do not bypass the initial warmup.
Cause: The VetTest analyzer is below the proper operating temperature.
Solution: Review Choosing the Analyzer Location, page B–1.
Very low temperatures can cause considerable damage to the
analyzer. Permanent damage may be caused if analyzer is switched
on at temperatures below freezing. Analyzer should be unpacked
BOEBMMPXFEUPDPNFUPSPPNUFNQFSBUVSF×o×$×o×'GPS
a minimum of one hour after it has been in a cold environment.

Table of Contents

Related product manuals