EasyManua.ls Logo

illumina NextSeq 500 - Chapter 4 Maintenance; Manual Wash Procedures

illumina NextSeq 500
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 Maintenance
Introduction 24
Perform a Manual Wash 24
Software Updates 26
Shut Down the Instrument 27
Introduction
Maintenance procedures include manual instrument washes and system software updates when available.
u Instrument washes—An automatic post-run wash after each sequencing run maintains instrument
performance. However, a manual wash is required periodically under certain conditions. See
Perform a
Manual Wash
on page 24.
u Software updates—When an updated version of the system software is available, you can install the
update automatically through a connection to BaseSpace or manually after downloading the installer from
the Illumina website. See
Software Updates
on page 26.
Preventive Maintenance
Illumina recommends that you schedule a preventive maintenance service each year. If you are not under a
service contract, contact your Territory Account Manager or Illumina Technical Support to arrange for a
billable preventive maintenance service.
Perform a Manual Wash
Manual washes are initiated from the Home screen. Wash options include the Quick Wash and the Manual
Post-Run Wash.
Wash Types Description
Quick Wash
Duration: 20 minutes
Flushes the system with a user-supplied wash solution of laboratory-grade water and
Tween 20 (buffer wash cartridge).
Required every 14 days that the instrument is idle or after a shutdown.
Manual Post-Run Wash
Duration: 90 minutes
Flushes the system with a user-supplied wash solution of laboratory-grade water and
Tween 20 (buffer wash cartridge) and 0.12% sodium hypochlorite(reagent wash
cartridge).
Required if the automatic post-run wash was not performed.
A manual wash requires the reagent wash cartridge and buffer wash cartridge provided with the instrument,
and a used flow cell. A used flow cell can be used up to 20 times for instrument washes.
Figure 19 Reagent Wash Cartridge and Buffer Wash Cartridge
Document # 15046563 v04
For Research Use Only. Not for use in diagnostic procedures.
24

Related product manuals