INFICON Operation | 6
XL3000flex-Operating-instructions-jina83en1-07-(2105) 35 / 108
6.2.2 Setting date and time
Setting the date
Format: DD.MM.YY
Control unit Main Menu > Settings > Date/Time > Date
LD protocol Command 450
ASCII protocol *HOUR:DATE
Setting the time
Format: hh: mm
Control unit Main Menu > Settings > Date/Time > Time
LD protocol Command 450
ASCII protocol *HOUR:TIME
6.2.3 Selecting a unit for the leak rate
Leak rate unit display
Selecting the leak rate unit in the display for sniffing
0 mbar l/s (factory setting)
1 Pa m
3
/s
2 atm cc/s
3 Torr l/s
4 ppm
5 g/a
6 oz/yr
7 sccm
8 sft
3
/yr
Control unit Main Menu > Display > Units (Display) > Leak Rate Unit SNIF
LD protocol Command 396 (sniffing)
ASCII protocol Command *CONFig:UNIT:SNDisplay
Leak rate unit interface
Selecting the leak rate unit of the interfaces for sniffing
0 mbar l/s (factory setting)
1 Pa m
3
/s
2 atm cc/s
3 Torr l/s
4 ppm