Ambient conditions
XL3000flex, XL3000flexRC
Permissible ambient temperature (during
operation)
10 °C ... 40°C
Max. altitude above sea level 2000 m
Max. relative humidity ≤ 31°C 80 %
Max. relative humidity > 31°C 50 %
Storage temperature -20 °C ... 60°C
Degree of contamination 2
4.8 Factory settings
The following table shows the factory settings in the "Sniffing" mode.
Parameter Factory setting
AO upper limit exp. 1 x 10
-5
Equivalent factor mass 2 (H2) 1.0
Equivalent factor mass 3 1.0
Equivalent factor mass 4 (He) 1.0
Molar mass of the equivalence gas (test gas
mass 2 (H2))
2.0
Molar mass of the equivalence gas (test gas
mass 3)
3.0
Molar mass of the equivalence gas (test gas
mass 4 (He))
4.0
Operation mode XL Sniffer Adapter
Bus module address 126
Pressure capillary blocked
(Low Flow) 0.2 mbar
Pressure capillary broken
(Low Flow) 0.6 mbar
Clogged pressure capillary monitoring
- with XL Sniffer Adapter (High Flow) 150 mbar
Broken pressure capillary monitoring
- with XL Sniffer Adapter (High Flow) 400 mbar
Pressure unit (interface) mbar
Emission On
Filter leak rate threshold 1 x 10
-10

Table of Contents

Other manuals for Inficon XL3000flexRC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon XL3000flexRC and is the answer not in the manual?

Inficon XL3000flexRC Specifications

General IconGeneral
BrandInficon
ModelXL3000flexRC
CategorySecurity Sensors
LanguageEnglish

Related product manuals