6.3.4.6 Automate external calibration with CalMate (optional)
WARNING
Danger to pacemaker wearers from magnets
The calibration adapter contains magnets with which it adheres to the calibration leak.
If you wear a pacemaker, do not install it yourself.
If you wear a pacemaker, always maintain a distance of at least 10 cm from the
calibration adapter when operating the device.
In sniffer mode, it is possible to automate calibrations with an external calibration leak.
ü The CalMate calibration adapter is mounted on an external INFICON sniffer
calibration leak so that the opening in the adapter is directly above the outlet of the
calibration leak.
ü The calibration adapter is connected to the leak detector by cable. For further
details, refer to the "CalMate Calibration Adapter for Test Leaks" operating
instructions.
ü If there is no separate interface on the back of the XL3000flex for connecting the
connection cable, you have established the connection via the IO1000 module. See
also "Device setup [}14]".
ü Using the CalMate on the XL3000flex requires a basic unit software version of
V2.74 or higher.
ü In the sniffer leak detector, the gas type (mass) and leak rate of the external
calibration leak are set.
1
Start the calibration check in the XL3000flex by inserting the sniffer tip into the
calibration port on the CalMate.
ð A light barrier in the CalMate adapter detects when a sniffer tip is held in the
calibration opening.
ð If a significant error is detected during this check, the leak detector
recommends another calibration.