INFICON Operation | 6
XL3000flex-Operating-instructions-jina83en1-07-(2105) 61 / 108
Control unit Main Menu > Functions > CAL > Settings > CAL request. >
Calibration request
or
Main Menu > Settings > Set-up> CAL request. > Calibration
request
LD protocol Command 419
ASCII protocol *CONFig:CALREQ (ON,OFF)
Calibration warning
Wrn650
The warning message Wrn650 "Calibration within the first 20 minutes is not
recommended" can be allowed or suppressed.
0 OFF (suppressed)
1 ON (allowed)
Control unit Functions > CAL > Settings > CAL request. > Calibration
warning W650
or
Settings > Set-up> CAL request. > Calibration warning
W650
LD protocol Command 429
ASCII protocol *CONFig:CALWarn ON (OFF)
6.3.4.2 External Calibration Configuration and Start
Prerequisite for the calibration with the internal calibration leak is the one-time entry of
the leak rate of the calibration leak.
In Sniffer mode, sniffing with the sniffer line is always performed on the open
calibration leak.
Leak rate of external
calibration leak sniffing
Define the leak rate of the calibration leak you wish to use during calibration.
Calibration will not be possible unless you enter the value here.
A specific leak rate must be set for each gas (mass).
Control unit Main Menu > Settings > Set up > Operation modes > Sniffing
> Ext. calibration leak > Mass 2 (3, 4)
or
Main menu > Functions > CAL > Settings > Ext. calibration
leak (for current mass in selected unit)
LD protocol Command 392