6.2.21 Assigning analog outputs of the I/O module
The analog outputs of I/O module IO1000 can with assigned with different
measurement value displays.
Possible functions: see the following table
Control unit Main Menu > Settings > Set up > Interfaces > I/O module >
Analog outp. > Config. Analog outputs 1/2
LD protocol Commands 222, 223, 224
ASCII protocol Command *CONFig:RECorder:LINK1
Command *CONFig:RECorder:LINK2
Command *CONFig:RECorder:SCALE
Command *CONFig:RECorder:UPPEREXP
Limit values can be defined for the output voltages.
SNIF: Min. 1 x 10
-9
...1 x 10
-1
mbar l/s
Max. 1 x 10
-8
...1 x 10
-1
mbar l/s
Control unit Main Menu > Settings > Set up > Interfaces > LR limits
LD protocol Command 227 (Snif)
ASCII protocol Command *CONFig:LIMITS:SNIF
Functions, assignment of analog outputs:
Off The analog outputs are switched off
(Output voltage = 0 V).
Pressure p1 / Pressure p2 1 ... 10 V; 0.5 V / decade;
1 V = 1 x 10
-3
mbar
Leak rate mantissa 1 ... 10 V; linear; in the selected unit Useful only if the other analog
output is assigned "Leak rate
exponent".
Leak rate exponent 1 ... 10 V; 0.5 V / decade;
Step function;
1 V = 1 x 10
-12
; in selected unit
Useful only if the other analog
output is assigned “Leak rate
mantissa” or “Leak rate ma. Hys."
is occupied.
Linear leak rate x... 10 V; linear;
in the selected unit