INFICON Operation | 6
XL3000flex-Operating-instructions-jina83en1-07-(2105) 37 / 108
Function of right sniffer
key
Activate or deactivating the right key of the SL3000XL sniffer line (switching between
low flow and high flow). Deactivating the key prevents an inadvertent influencing of
the measurement.
Control unit Settings > Set up > Operation modes > Sniff > Sniffer > Keys >
Sniffer flow key
LD protocol Command 415
ASCII protocol Command *CONFig:HFButton
Search Function
When the search function is activated, the alarm is automatically connected to
Trigger 2 as soon as it is switched to High Flow.
Switched-off Search Function: Alarm, when Trigger 1 is exceeded.
Switched-on Search Function and operation in Low Flow: Alarm, when Trigger 1
is exceeded.
Switched-on Search Function and operation in High Flow: Alarm, when Trigger 2
is exceeded.
0 Off
1 On
Control unit Setting > Trigger > Search
LD protocol Command 380
ASCII protocol Command *CONFig:SEARch
In the SL3000XL the following are dependent on the trigger used; the leak rate bar,
changing the background lighting, the beeper and changing the sniffer tip lighting.
Sniffer LEDs:
Brightness
Set the brightness of the LEDs designed to illuminate the spot under examination.
This setting refers to the measurement process without LED alarm configuration, see
below.
From "0" (off) to "6" (max.)
Control unit Settings > Set up > Operation modes > Sniff > Sniffer > LED >
Sniffer LED brightness
LD protocol Command 414
ASCII protocol Command *CONFig:BRIGHTness
Sniffer LEDs: Alarm
configuration
Behavior of the LEDs on the sniffer, when trigger value 1 is exceeded.
Off No response
Flashing The LEDs are flashing
Brighter The LEDs shine with maximum brightness.