LDS3000
Error number Limit values Cause
LDS1000
Protocol
Binary or
ASCII
protocol
compatibilit
y mode
LDS1000/
LDS2010
Err306 Anode voltage faulty 36 127 40 V
deviation
from the
default
value
The anode voltage does not
match the default value or the set
value is outside the allowable
setting range.
Wrn310 Cathode 1 is defective 45 136 Cathode defective, line to cathode
interrupted, IF board or MSB
defective
Wrn311 Cathode 2 is defective 46 137 Cathode defective, line to cathode
interrupted, IF board or MSB
defective
Err312 Cathode defective 47 138 Cathode defective, line to cathode
interrupted, IF board or MSB
defective
Err340 Emission error 44 135 < 90% of
the target
value
> 110% of
the target
value
Emission was stable previously,
pressure probably too high,
message after 15s
Wrn342 Cathode not connected 47 138 Both cathodes defective during
self-testing after switch on or plug
not connected
Wrn350 Suppressor not connected 39 130 Supressor cable is not plugged in
or defective during the self-test
after switching on
Wrn352 Preamplifier not connected Preamplifier defective, cable not
plugged in
Err358 Preamplifier oscillates
between 2 ranges
Signal varies too much (see
command 1120)
Preamplifier defective
Wrn359 Overdriven preamplifier 31 123 Signal too large preamplifier
defective