Optimas System Operator Manual
Section 4: Device Description
Figure 4-10: Fractora Treatment Screen
The Fractora Treatment Screen contains:
The frame selects parameters that can be changed by the function
keys.
Allows selecting between different tip types of Fractora.
Delivered energy is changed from level 5 to 62 energy levels and
the System starts up at the minimal energy setting.
Allows choosing between single pulse and pulse trains ranging from
0.5 to 2 pulses per second (pps).
The Counter can be reset.
Shows number of pulses delivered on one zone and the total
number of pulses from the beginning of the treatment.
The System has three treatment modes: Standby, Ready, and
Active.
Standby mode- allows the user to set treatment parameters.
Activation of energy is not allowed in Standby mode.
Ready mode - the System is waiting for a signal from the foot
switch to activate the energy. Any attempt to change the
treatment settings switches the System to Standby mode.
Active Mode - when the signal from the Footswitch is indicated,
the System switches to Active mode. Any attempt to change the
treatment settings switches the System to Standby mode.
Return to the Main Menu to select another applicator and change
the applicator if needed.
Figure 4-11: Forma Treatment Screen