Optimas System Operator Manual
Section 4: Device Description
The Lumecca 580 or 515nmTreatment Screen contains:
The frame selects parameters that can be changed by the function
keys.
Fluence (light energy density) is changed within the limits from 5 to
30J/cm² for Lumecca and the System starts up at the minimal
setting.
Normal and Strong cooling can be selected for Lumecca.
Short or Long pulse can be selected.
Repeat mode in Normal or Fast repetition rate or in Off as a single
pulse mode.
The Pulse Counter can be reset.
Shows the number of pulses delivered from the beginning of the
treatment.
The System has three treatment modes, Standby, Ready, and
Active.
Standby mode - allows the user to set treatment parameters.
Activation of energy is not allowed in Standby mode.
Ready mode - the System is waiting for a signal from the
Footswitch to activate the energy. Any attempt to change the
treatment settings switches the System to Standby mode.
Active mode - when the signal from the Footswitch is indicated,
the System switches to Active mode. Any attempt to change the
treatment settings switches the System to Standby mode
Return to the Main Menu to select another applicator and change
the applicator if needed.