Section 4: Device Description
27 of 87
The Lumecca 580 or 515nmTreatment Screen contains:
Selection
Frame
The frame selects parameters that can be changed by the function
keys.
Fluence
Fluence (light energy density) is changed within the limits from 5 to
30J/cm² for Lumecca and the System starts up at the minimal
setting.
Cooling
Normal and Strong cooling can be selected for Lumecca.
Pulse Width
Short or Long pulse can be selected.
Pulse Mode
Repeat mode in Normal or Fast repetition rate or in Off as a single
pulse mode.
Counter Reset
The Pulse Counter can be reset.
Pulse Counter
Shows the number of pulses delivered from the beginning of the
treatment.
System Mode
The System has three treatment modes, Standby, Ready, and
Active.
Standby mode - allows the user to set treatment parameters.
Activation of energy is not allowed in Standby mode.
Ready mode - the System is waiting for a signal from the
Footswitch to activate the energy. Any attempt to change the
treatment settings switches the System to Standby mode.
Active mode - when the signal from the Footswitch is indicated,
the System switches to Active mode. Any attempt to change the
treatment settings switches the System to Standby mode
Main Menu
Return to the Main Menu to select another applicator and change
the applicator if needed.