Working with ENG: EO425 Page 33
ENG/DLB Instruction for Use - English
Date: 2011-09-27 Page 33
Figure 1.25 Compensating for drift in saccade test
This will adjust the tracings and re-center them:
The longer the test is run the more likely it is that you will experience signal drift.
Accommodating for Noisy Recordings
In certain situations, it is possible that there may be noise or artifact in your ENG recordings that
the software may misinterpret and mark as beats of nystagmus.
Figure 1.26 Noise in ENG recording
To accommodate for this noise so that fewer nystagmus beats are incorrectly marked, you can go
into configuration for the specific test by double-clicking in the test window. Go to the Analysis tab
and increase the minimum amplitude of fast phase to 2 or 3 degrees.
Figure 1.27 Configure Test Analysis Tab