Invacare®Robin™
5Troubleshooting
5.1TroubleshootingTable
SYMPTOMSPROBLEMSOLUTION
Emergencystopactivated
Resettheemergencystop.(Refertouser
manual)
Systemdoesnothavepowerorbatteryis
discharged
Chargethebatteryorusetheemergency
battery.Referto4.1ChargingtheBattery,page
20
.
Ceilinghoistdoesnotrespondtothehand
control’sbuttonsandup/downbuttonsonthe
ceilinghoistarenotresponding.
Fuseisblown
Replacethefuse.Referto3.8Replacingthe
fuse,page15
.
Handcontrolnotconnectedproperly
Checkthatthehandcontrolcableisproperly
pluggedintothehoist.Referto3.3Replacing
thehandcontrol,page12
.
Ceilinghoistdoesnotrespondtothehand
control’sbuttons,butup/downbuttonsonthe
ceilinghoistareresponding.
HandcontroldefectiveReplacethehandcontrol.Referto3.3Replacing
thehandcontrol,page12
.
Batterynearlydischarged
Chargethebatteryorusetheemergency
battery.Referto4.1ChargingtheBattery,page
20
.
Liftingstrapswillonlymovedownandnotup.
LoadistoohighReducetheload.Thehoistisequippedwithan
over-currentsafetysystem.
HandcontroldefectiveReplacethehandcontrol.Referto3.3Replacing
thehandcontrol,page12
.
Liftingstrapswillonlymoveupandnotdown.
TheloosestrapfunctionactivatedChecktheloosestrapfunction.Referto5.3
CheckingLooseStrapFunction,page26.
Ceilinghoistspeedisthesamewithandwithout
load.
Non-loadquickmovementfunctionactivated
permanently
Checkthequickmovementswitch.Referto5.4
CheckingTwinSpeedFunctionality,page26
.
LiftinghooksdonotturnfreelyenoughCleanthehooksorreplacethestraps. Liftingstrapsaretwistedanddonotstraighten
out.
StrapsarefrayedReplacethestraps.Referto3.5ReplacingStrap
andHookAssemblies,page12.
24
1586747-A