Invacare®Robin™
5.2Indicatorlightandaudiosignal
Hoist
Status
IndicatorlightAudiosignal
Off/Standby
Hoisttraversing(Robin
Mover™only)
Off
No
Hoistinuse(Up/Down)
BlueNo
LowbatteryBlueFlashingYes
Theblueindicatorlightontheundersideofthehoistturnsoff
automaticallyafter30seconds.
WallMountedCharger
Status
IndicatorlightDuration
BatterynotconnectedYellow
n/a
Initializingandanalysis
ofbattery
Yellow
12–14seconds
FastchargeRed
5min—3hours
Charged(trickle
charge)
Greenn/a
Faultcondition(battery
defect)
Alternatingred
andgreen
n/a
5.3CheckingLooseStrapFunction
Acorrectlyadjustedloosestrapswitchensuresthatthestrapsstop
loweringwhentheytouchanobstruction.
1.Withtheceilinghoistremainingonthetrack,checkthereare
noobstructionsundereitherstraphook.
2.Placeanobstructionunderoneofthestraphooks,thenpress
andholdthedownbuttononthehandcontrol,
Expectedresult:Whenthestraphooktouchestheobstruction,
thestrapstopslowering.
3.IfthestrapdoesNOTstoplowering,adjusttheloosestrap
switch.Refertochapter3.10AdjustingLooseStrapSwitch,page
16
.
4.Repeatsteps2and3forthesecondstrap.
5.4CheckingTwinSpeedFunctionality
1.Withthehoistonthetrackandbothstrapsfreefrom
obstructions,pressandholdtheupbuttonwithnoloadapplied.
Notespeed(40mm/s)andsoundofmotor.
2.Whilstcontinuingtoholdtheupbutton,applyaloadofatleast
10kg(e.g.bypullingdownonbothstraps).
Notethespeedreduction(30mm/s)andmotorsoundchange.
3.Ifthereisnospeedreduction,replacethetwinspeedswitch.
Refertochapter3.11ReplacingtheTwinSpeedSwitch,page18.
261586747-A