Technicaldata
6.5Electromagneticcompliance(EMC)
information
MedicalElectricalEquipmentneedstobeinstalledandusedaccording
totheEMCinformationinthismanual.
ThisequipmenthasbeentestedandfoundtocomplywithEMClimits
specifiedbyIEC/EN60601-1-2forClassBequipment.
PortableandmobileRFcommunicationsequipmentcanaffectthe
operationofthisequipment.
Otherdevicesmayexperienceinterferencefromeventhelowlevels
ofelectromagneticemissionspermittedbytheabovestandard.To
determineiftheemissionfromtheliftiscausingtheinterference,run
andstoprunningthelift.Iftheinterferencewiththeotherdevice
operationstops,thentheliftiscausingtheinterference.Insuchrare
cases,interferencemaybereducedorcorrectedbythefollowing:
•Reposition,relocate,orincreasetheseparationbetweenthe
devices.
6.5.1Guidanceandmanufacturer’sdeclaration
Test
Basic
Standard
ComplianceLevel/
FrequencyRange
MainsPort
Conducted
Emissions
EN
55011:2007
Group1ClassB
Radiated
Emissions
EN
55011:2007
Group1ClassB
MainsHarmonics
EN61000-
3-2:2006+
A1:2009+
A2:2009
ClassA
Test
Basic
Standard
ComplianceLevel/
FrequencyRange
Voltage
Fluctuations
EN61000-3-
3:2008
dc%/dmax/d(t)/Pst
Electrostatic
Discharge
EN61000-4-
2:1995
±2/4/6kVContact,
±2/4/8kVAir
RadiatedField
Immunity
EN61000-4-
3:2006
3V/m,80%1kHzAM,
80MHz-2.5GHz
EFT/Bursts
EN61000-4-
4:2004
±2kVL/N/E/LNE
SurgeImmunity
EN61000-4-
5:2006
±0.5/1kVL-N,±0.5/1/2kV
L-E/N-E
ConductedRF
Immunity
EN61000-4-
6:2007
3V,80%1kHzAM,0.15MHz
-80MHz
MagneticField
Immunity
EN61000-4-
8:1993
3A/m,50Hz&60Hz
Dips&
Interruptions
EN61000-4-
11:2004
-95%/-60%/-30%