Invacare®Robin™
6.3ElectricalSystem
Robin™RobinMover™
Maximumcurrent
input
0.9A
Protectionclass
1
Entiredevice:IPx4
(ceilinghoist:IP24;handcontrol:IP44)
Insulationclass
ClassIIequipment
TypeBappliedpart
Appliedpartcomplyingwiththespecified
requirementsforprotectionagainstelectrical
shockaccordingtoIEC60601-1.
Soundlevel50–55dB(A)
Numberofliftsper
charge
65liftsof0.5mwith80kg
Intermittent
(periodicmotor
operation)
10%,max.2min./18min.
Battery
NiMHBattery24V/2.0Ah
ChargingtimeUpto3hoursat20°C
Liftingspeedwith
200kgload
2.1cm/s2.1cm/s
Traversespeedin
thetracks
n/a14cm/s
Charger
Voltageoutput
charger
24VDC,max.250VA
Voltagesupply
charger
230VAC,50Hz
Protectionclass
1
charger
IP41orIP67(dependingonversion)
1
Seeproductlabelandlabeloneachelectricdeviceforcorrect
protectionclass.ThelowestIPclassificationdecidestheoverall
classificationofthedevice.
•IPx4:Protectedagainstwatersplashedfromanydirection.
•IP24:Protectedagainstobjectslargerthan12.5mmand
protectedagainstwatersplashedfromanydirection.
•IP44:Protectedagainstobjectslargerthan1mmandprotected
againstwatersplashedfromanydirection.
•IP41:Protectedagainstobjectslargerthan1mmandprotected
againstverticallydrippingwater.
•IP67:Dust-tightandprotectedagainsttemporaryimmersion
intowater.
6.4Environmentalconditions
Storageand
transportation
Operation
Temperature
-30°Cto+50°C+5°Cto+40°C
Relativehumidity
10%to70%
20%to90%at30
°C,notcondensing
Atmospheric
pressure
700hPato1060hPa
301586747-A