SPECIFICATIONS
Specifications for the HEMOCHRON Signature Elite Whole Blood Microcoagulation System are listed below.
Dimensions and Weight
≤ 10% C.V. for whole blood samples
Full-Charge Operating Time
49 test cycles (at 150 sec per test)
17 test cycles (> 500 sec per test)
100 to 240 VAC, 50 to 60 Hz
12.0 VDC, 3.5 Amps (42 Watts, 144 BTU/hr) minimum
Supported Barcode Formats
UPC/EAN, Code 128, Code 39, Trioptic Code 39,
Code 93, Interleaved 2 of 5, Discrete 2 of 5, Codabar,
and MSI Plessey
IEC 60825-1 class 1 or class 2 laser product complies with
21 CFR 1040.10 and 1040.11 except for deviations pursuant
to Laser Notice No. 50 dated June 24, 2007.
Class 1: 630-670 nm
Class 2: 630-670 nm
Class 1: 0.7mW @peak
Class 2: 1.5 -1.9 mW @peak
Calibration
The Signature Elite instrument is calibrated at the manufacturing facility to test and verify all functions. The
instrument is also self-calibrating, as all instrument functions are continuously monitored and verified by the
instrument software when a test is performed. The instrument does not require additional calibration by the
user.
For Class 2 Laser Instruments:
CAUTION: CLASS 2 LASER RADIATION WHEN OPEN. DO NOT STARE INTO THE BEAM.
CAUTION: This manual provides safety instructions for instruments equipped with class 1 or class 2
barcode scanner lasers. Please refer to the back panel of the instrument to identify the class of laser
and follow appropriate instructions provided in this manual.