Acronyms for Error Messages
The following table lists the acronyms for Error Messages when viewing stored Patient or QC records:
Acronym Description Display Message
ORH Out of Range High Out of Range-Hi
ORL Out of Range Low Out of Range-Lo
SNS Sample Not Seen Sample Not Seen
STS Sample Too Small Sample Too Small
STL Sample Too Large Sample Too Large
TTO Temp Timeout Heater Too Cool
TTH Temp Too High Heater Too Hot
SPF Sample Position Fault Sample Pos Fault
UAT User Aborted Test User Abort
BAT Battery Fault Battery Fault
PMS Premature Sample Premature Sample
DEF Detector Fault Detector Fault
BTH Battery Too High Battery Too High
XTH External Power Too High External Too High
RTC Real Time Clock Fault RTC Fault