3. Pre-stored only: Enter a check mark to specify the cuvette information can only be selected from a
list of lot numbers specified by the supervisor.
Note: Up to five cuvette lot numbers and expiration dates can be specified for each assay (see
page 64).
Note: If more than five cuvette lot numbers per assay are scanned or entered, the lot number(s)
with the closest expiration date will be deleted.
QC Lot Options
1. Required: Enter a check mark to require entry of a QC lot number when running a quality control
test.
2. Scan only: Enter a check mark to specify the QC lot number can only be entered using the barcode
scanner.
Note: If Scan only is enabled, the operator is prompted to scan the QC information when
running a test. The operator can still enter the QC information manually by pressing the 7 key
(as indicated by the 7=Manual Entry prompt), but the QC information must be entered twice.
Refer to page 33 for additional information.
3. Pre-stored only: Enter a check mark to specify the QC lot number can only be selected from a stored
list.
Note: Up to five QC lot numbers, expiration dates, and ranges can be specified for each assay
for both level 1 and level 2 controls (see page 65).
Note: If more than five QC lot numbers per assay are scanned or entered for level 1 and/or level
2 controls, the lot number(s) with the closest expiration date will be deleted.
Saving the Options
1. Click Save & Exit to save the specified options and cancel display of the dialog box.
Note: Cancel clears all fields without saving any parameters.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITC HEMOCHRON Signature Elite and is the answer not in the manual?

ITC HEMOCHRON Signature Elite Specifications

General IconGeneral
BrandITC
ModelHEMOCHRON Signature Elite
CategoryMedical Equipment
LanguageEnglish

Related product manuals