Note: Subsequent dates and times will be determined automatically from the QC interval.
6. Auto EQC: Enter a check mark to specify that EQC should be automatically run at the specified
interval.
Note: It Auto EQC is selected and the HEMOCHRON Signature Elite is connected to a power
adapter, the instrument will remain on and perform the required EQC test at the appropriate
interval. The results will be automatically stored in the database.
Important: It Auto EQC is selected, ensure that cuvettes are routinely removed from the
instrument after completing a test. If EQC is run while a cuvette is inserted, the message
Detector Blocked is displayed and the EQC test fails.
7. Click Save to save the EQC lockout parameters.
Note: Click Cancel to clear the fields without saving the modified parameters. Save & Exit saves
the EQC lockout parameters and closes the dialog box.
To Specify QC Exceptions:
1. Click Lock-Out Control on the Configuration Module dialog box (page 53) to display the QC
Lockout dialog box.
2. Select the QC Exceptions tab:
Note: Entries made on the QC Exceptions tab apply to all assays.
911 Limit
1. Enter the maximum number of patient tests (the 911 Limit, from 0 to 240 tests) that can be run
after the specified EQC/LQC lockout interval has been reached an assay. When the 911 limit is
reached for an assay, that assay cannot be run until a successful QC test is run for that assay or until
the instrument is unlocked (see page 61) using Configuration Manager.
Note: The number of patient tests that have already been run against the current 911 limit is
displayed when read from an instrument.
Failure Limit
1. EQC Failure Limit: Enter the maximum number of times that a failed EQC test can be rerun before
the system will lock (from 1 to 9 tests).
2. LQC Failure Limit: Enter the maximum number of times that a failed LQC test can be rerun before
the system will lock (from 1 to 9 tests).
Note: Setting the failure limit to 0 (zero) disables lockout by failure count.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITC HEMOCHRON Signature Elite and is the answer not in the manual?

ITC HEMOCHRON Signature Elite Specifications

General IconGeneral
BrandITC
ModelHEMOCHRON Signature Elite
CategoryMedical Equipment
LanguageEnglish

Related product manuals