Scanning Lot Information for Cuvettes or Quality Control Materials
1. On the Supervisor menu, press 1 to display the Scan Lot screen:
2. Hold the instrument over the barcode on the cuvette wrapper or QC package:
Note: Position the scanner port of the instrument approximately four inches from the
barcode (see page 5).
3. Press PRINT/SCAN to scan the barcode label. The scanned lot number and a Stored confirmation
message are displayed, and the entry is saved.
Note: Additional lots can be scanned by repeating Steps 2 and 3.
4. Press CANCEL to exit.
Scanning Operator IDs
1. On the Supervisor menu, press 2 to display the Scan OID screen:
2. Position the scanner port of the instrument approximately four inches from the barcode
(see page 5).
3. Press PRINT/SCAN to scan the barcode. The scanned Operator ID and a Stored confirmation
message are displayed, and the entry is saved.
Note: Additional Operator IDs can be scanned by repeating Steps 2 and 3.
4. Press CANCEL to exit.
Setting the System Time or System Date
For convenience, the Supervisor menu includes the same commands that are on the main menu for
changing the system time and system date.
1. On the Supervisor menu, press 3 to set the system time.
2. On the Supervisor menu, press 4 to set the system date.
Note: Procedures for setting the system time and system date are described on page 15.
(cuvette wrapper)
barcode
(QC package)
barcode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITC HEMOCHRON Signature Elite and is the answer not in the manual?

ITC HEMOCHRON Signature Elite Specifications

General IconGeneral
BrandITC
ModelHEMOCHRON Signature Elite
CategoryMedical Equipment
LanguageEnglish

Related product manuals