Load Function
2. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
3. Select 1. LIST: Off and press [Enter].
4. Press the Left/Right key to select Open and press [Enter].
5. Press the Left/Right key to select Internal and press [Enter].
6. Set the name of the List file to be recalled (that is, the file name set in Edit),
and press [Enter].
The interface returns to the main interface of the LIST function, which is dis-
played as follows:
FUNCTION LIST
Run Open Edit Export
7. Press the Left/Right key to select Export and press [Enter].
8. Press the Left/Right key to select Yes and press [Enter].
Indicates that the selected List file in Open is exported to the USB flash
drive.
Run the List Program
You can select a List file to execute according to your needs, so that the instru-
ment inputs the corresponding waveform sequence. Take the List file in Internal
as an example. The steps are as follows:
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
2. Select 1. LIST: Off and press [Enter].
3. Press the Left/Right key to select Open and press [Enter].
4. Press the Left/Right key to select Internal and press [Enter].
5. Set the name of the List file to be recalled (that is, the file name set in Edit),
and press [Enter].
The interface returns to the main interface of the LIST function, which is dis-
played as follows:
FUNCTION LIST
Run Open Edit Export
6. Press the Left/Right key to select Run and press [Enter].
At this point, the interface returns to the main interface of the system and
Lxx/xxxx WTG is displayed in the lower right corner.
7. Turn on the [On/Off].
8. Based on the selected trigger method, perform the trigger operation.
Take the manual trigger as an example. Press [Shift]+[On/Off](Trigger) on
the front panel to run the selected List file. For details about the List trigger
method, see 5.8 Select Trigger Source (Trig Source).