Basic Operation
• Voltage and current
Records current and voltage data during the data acquisition period.
Configure the Function Menu
1. Press the composite button [Shift]+[1] (Log) on the front panel to enter the
configuration menu of the data logging function.
The parameters are described as follows:
Menu Item Description Setting
Sample
Period
This parameter indicates the data
sampling interval (unit: s), that is, the
test data is recorded once every X
seconds.
The input range supported by the in-
strument is: 0 - 9
Set the value by in-
putting the value
through the number
key or rotating the
knob.
Duration This parameter indicates the time for
data recording (unit: s), namely, the
data recording will be completed in Y
seconds and be ready for the next
data record.
The input range supported by the in-
strument is: 0 - 3600
Source This parameter indicates the source
of recorded data, including voltage
(V), current (I), voltage and current
(V/I).
Use the Left and
Right arrow key or
the rotation knob to
select.
Data Type This parameter indicates the type of
recorded data, including:
• Average: This is the default se-
lected data type, namely, the
average of all data acquired in the
data recording period as indicated
in the table of saved data.
• Max/Min: If this option is se-
lected, it means the maximum
value and minimum value of data
acquired in the data recording pe-
riod as indicated in the data sav-
ing table.
Copyright © Itech Electronic Co., Ltd.