EasyManua.ls Logo

Ivoclar Vivadent Programat CS6 - Page 61

Ivoclar Vivadent Programat CS6
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61
4. What If...
4136 Yes The firing chamber temperature is higher than the maximum permissible
start temperature. Please wait until the temperature has dropped below the
maximum permissible start temperature. For more detailed information,
please refer to the Operating Instructions.
4137 Yes The tightness of the vacuum system is not given. Check the sealing
surfaces for possible contamination.
4138 Yes The set vacuum end value could not be achieved. Check the vacuum
pump.
4146 No The self test could not be performed due to a power failure during a
program in progress. Please restart the furnace.
4148 No An error has occurred while operating the motor. Restart the furnace. If this
error occurs again, please contact your Service Centre!
4149 No Self-test not successful. Restart furnace and maybe contact After Sales
Service.
* Furnace head opens if this error occurs!
** A running program will be aborted
*** Error cannot be acknowledged – programs cannot be started
Please contact the Ivoclar Vivadent After Sales Service, if one of the following error numbers is being displayed:
25, 29, 43, 44, 45, 46, 47, 48, 56, 103, 107, 108, 143, 144, 145, 146, 147, 148, 530, 531, 532, 700, 701, 707, 1010,
1011, 1012, 1013, 1014, 1015, 1016, 1017, 1018, 1019, 1024, 1025, 1026, 1028, 1029, 1030, 1031, 1032, 1033, 1034,
1207, 1300, 1301, 1303, 1304, 1305, 1310, 1401, 1402, 1500, 1652, 1750, 1751, 1752, 1753, 2001, 2002, 2750, 2751,
2752, 2753, 2770, 3000, 3001, 3002, 3010, 3011, 4001, 4049, 4060, 4061, 4062, 4063, 4064, 4065, 4066, 4067, 4080,
4140, 4141, 4142, 4143, 4144, 4145