60
4. What If...
4050
**,***
No The furnace head cannot be moved to its end position. The furnace head
might be blocked by external mechanical influences! If this is not the case,
please contact your Service Centre!
"4051
**"
No The maximum heating time with the furnace head open was exceeded. The
process is aborted for safety reasons. If this error occurs repeatedly, please
contact your Service Centre.
4070 No The maximum permissible housing temperature was exceeded. Check to
make sure that all air vents are unobstructed and that the fans are not
blocked. If this error occurs repeatedly, please contact your Service Centre.
4071 Yes The temperature inside the furnace is too low for operation. Wait until the
furnace has reached room temperature.
4072 No The temperature sensor monitoring the housing temperature cannot be
read. A program start is, therefore, not possible. Contact your Service
Centre.
4100 No PowerUnit: Unknown error Communication error, please contact your Service Centre.
Error code: ****
4101 No PowerUnit: I2C write error Communication error, please contact your Service Centre.
4102 No PowerUnit: I2C read error Communication error, please contact your Service Centre.
4103 No PowerUnit: m_pI2CCommDriver
= NULL
Communication error, please contact your Service Centre.
4104 No m_pI2CCommDriver->is
DeviceDriverOpen() = false
Communication error, please contact your Service Centre.
4110 No PowerUnit: pCommunicationData
= NULL
Communication error, please contact your Service Centre.
4111 No PowerUnit: Requestqueue is full Communication error, please contact your Service Centre.
4112 No PowerUnit: No communication
with the PowerUnit possible
Communication error, please contact your Service Centre.
4120 No PowerUnit: Overvoltage
shutdown
An error has occurred with the heating system. Restart the furnace. If this
error occurs repeatedly, please contact your Service Centre.
4121 No PowerUnit: Overload shutdown An error has occurred with the heating system. Restart the furnace. If this
error occurs repeatedly, please contact your Service Centre.
4122 No PowerUnit: Internal temperature
is over 85°C, Power is shutdown
Excess temperature in the power pack. Switch off the furnace and let it
cool down. Make sure that the air supply is unobstructed. If this error
occurs again, please contact your Service Centre!
4123 No PowerUnit: Fan fail, Power is
shutdown
Cooling your furnaces does no longer work properly. Please switch off the
furnace and contact your Service Centre.
4124 No PowerUnit: Unit fail, Power is
shutdown
An error has occurred with the heating system. Restart the furnace. If this
error occurs repeatedly, please contact your Service Centre.
4125 No PowerUnit: Internal temperature
is over 75°C
Excess temperature in the power pack. Switch off the furnace and let it
cool down. Make sure that the air supply is unobstructed. If this error
occurs again, please contact your Service Centre.
4126 No PowerUnit: AC input < 180 Vac,
Output power down
The supply voltage is outside the acceptable range. Check the supply
voltage.
4127 No PowerUnit: AC input < 85 Vac,
Power is off
The supply voltage is outside the acceptable range. Check the supply
voltage.
4130 Yes The firing chamber temperature is higher than the maximum permissible
start temperature. The service position can only be reached once the
temperature in the firing chamber is lower than the maximum permissible
start temperature. For more detailed information, please refer to the
Operating Instructions.
4131 Yes The furnace head was not correctly swivelled sideways. Confirm the
message to move the motor into the service position.
4132 Yes The furnace head was not correctly swivelled to the centre.
4135 Yes The firing chamber temperature is higher than the maximum permissible
start temperature. Please wait until the temperature has dropped below the
maximum permissible start temperature. For more detailed information,
please refer to the Operating Instructions.