70
The furnace may only be transported as described in these Operating Instructions.
In order to avoid corrosion on the device and subsequent damage to the device, transport and store the furnace only
within the permissible temperature range and ambient conditions.
Acceptable operating conditions
Acceptable ambient temperature range: +5 °C to +40 °C (41°F to 104°F)
Acceptable humidity range: 80% maximum relative humidity for temperatures up to
31°C (87.8°F), gradually decreasing to 50 % relative
humidity at 40 °C (104°F); condensation excluded.
Acceptable altitude The furnace is tested for use at altitudes of up to 2000 m
above sea level.
Acceptable transportation and storage conditions
Acceptable temperature range –20 °C to +65 °C (-4 to 149°F)
Acceptable humidity range Max. 80 % relative humidity
Acceptable ambient pressure 500 mbar to 1060 mbar
Use only the original packaging together with the corresponding foam material for shipping purposes.
Remove the object tray from the firing chamber for transportation, pack it securely and ship it in the accessories
carton with the furnace.
The furnace may only be transported with the furnace head closed.
6 Notes on Storage and Transportation