9
2 Application
2.1 Installation and initial start-up
2.1.1 Unpacking and checking the contents
Remove the furnace components from their packaging and place them on a suitable table. Please observe the instructions
on the outer packaging.
There are no special transportation grips on the furnace. However, it can be carried by the grooves on the sides of the
furnace bottom. The furnace should be carried by two people. If this is not possible, the device should be carried with the
rear of the furnace facing the body for ergonomic reasons. The furnace must not be carried by holding on to the display or
the cooling tray. Check the delivery for completeness (see delivery form in Chapter 8) and transportation damage. If parts
are damaged or missing, contact your local Ivoclar Vivadent Service Centre.
We recommend keeping the original packaging for future service and transportation purposes.
2.1.2 Selecting the location
The device is only permitted for use in closed rooms within the specified ambient conditions (see Chapter 6).
Place the furnace on a flat non-combustible table using the rubber feet. Make sure that the furnace is not placed in the
immediate vicinity of heaters or other sources of heat. Make sure to observe a clearance between the wall and the rear of
the device as well as on the left side of at least 10cm for air circulation purposes. Furthermore, there must be sufficient
space above the furnace head so that the furnace head can open without difficulties. Also ensure that there is enough
space between the furnace and the user, as the furnace releases heat during the opening of the furnace head.
The furnace should neither be placed nor operated in areas where there is an explosion hazard.
2.1.3 Initial start-up
The start-up of the furnace is very easy and involves only few steps. Before you start, make sure that the voltage indicated
on the rating plate (32) complies with the local power supply. If this is not the case, the furnace must not be connected.
Mount the tong holder left or right on the back of the furnace.
Connections
– Power connection
Please make sure that the voltage indicated on the rating plate
complies with the local power supply. Subsequently, connect the power
cord (35) with the power socket of the furnace (33).
The device must only be operated with the supplied original power cord
or a replacement cord with the same specifications.
33
35