4. What If...
2201 Yes WLAN error. Please restart the furnace.
2749 Yes An error occurred while checking the infrared camera
2754 Yes Error self-test An error occurred while checking the infrared camera
2755 Yes Error self-test An error occurred while checking the infrared camera
2756 Yes Error self-test An error occurred while checking the infrared camera
2757 Yes Error self-test An error occurred while checking the infrared camera
2758 Yes Error self-test An error occurred while checking the infrared camera
2759 Yes Error self-test An error occurred while checking the infrared camera
3900
**
No Motorcontroller Error during communication with the motor of the lifting unit. If this error
occurs again, please contact your Service Centre!
3901
**
No Motorcontroller The motor for the lifting unit cannot be initialized. Restart the furnace. If
this error occurs again, please contact your Service Centre!
3902
**
No Motorcontroller Error during communication with the motor of the lifting unit. If this error
occurs again, please contact your Service Centre!
3903
**
No Motorcontroller Error during communication with the motor of the lifting unit. If this error
occurs again, please contact your Service Centre!
3904
**
No Motorcontroller Error during communication with the motor of the lifting unit. If this error
occurs again, please contact your Service Center!
3905
**
No "Motorcontroller
"
Error during communication with the motor of the lifting unit. If this error
occurs again, please contact your Service Centre!
3906
**
No Motorcontroller Error during communication with the motor of the lifting unit. If this error
occurs again, please contact your Service Centre!
3907
**
No Motorcontroller Error during communication with the motor of the lifting unit. If this error
occurs again, please contact your Service Centre!
3908
**
No Motorcontroller Error during communication with the motor of the lifting unit. If this error
occurs again, please contact your Service Centre!
3909
**
No Motorcontroller An error has occurred while operating the motor. Restart the furnace. If this
error occurs again, please contact your Service Centre!
3910
**
No An error has occurred while operating the motor. Restart the furnace. If this
error occurs again, please contact your Service Centre!
3911 No An error has occurred with the lifting unit. Please contact your Service
Centre. Error code: ****
4010 No PowerBoard The PowerBoard could not be initialized. The heater does not work. Restart
the furnace. If this error occurs again, please contact your Service Centre.
4011 No PowerBoard The heating relay could not be closed. The heater does not work.
4020 No WatchDog Internal software error. Restart the furnace. If this error occurs repeatedly,
please contact your Service Centre.
4030 Yes LifterPosition Open the furnace head for the program start.
4031 Yes T2 < T1 Enter a lower value for T1 or a higher value for T2.
4032 Yes T3 < T2 Enter a lower value for T2 or a higher value for T3.
4033 Yes T4 > max (T1, T2, T3) Enter a lower value for T4 than the maximum holding temperature.
4034 Yes T5 > T4 Enter a lower value for T5 than for T4.
4035 Yes V1 >= max (T1, T2, T3) Enter a lower value for V1 than the maximum holding temperature.
4036 Yes V2 >= max (T1, T2, T3) Enter a value for V2 that is lower than or equal to the maximum holding
temperature.
4037 Yes V1 > V2 Enter a lower value for V1 than for V2.
4038 Yes V1 = 0, V2 ! = 0 Enter a value for V1 that is higher than 0 or deactivate V2.
4039 Yes The firing chamber temperature is higher than the maximum permissible
start temperature. Please wait until the temperature has dropped below the
maximum permissible start temperature. For more detailed information,
please refer to the Operating Instructions.
4040 Yes Open the furnace head for the program start.
4041 Yes The firing chamber temperature is higher than the maximum permissible
start temperature for calibration programs. Please wait until the
temperature has dropped below the maximum permissible start
temperature. For more detailed information, please refer to the Operating
Instructions.

Other manuals for Ivoclar Vivadent Programat CS6